This section describes the requirements for chemiluminescent detectors (CLD)
used to measure NO (a) A CLD must meet the following requirements:
(1) Make sure your CLD meets the accuracy and precision specifications in ISO
8178-1 (incorporated by reference in ยง1065.1010).
(2) The NO to NO (3) Use an overflow sampling system for continuous CLDs. (In an overflow
system excess zero gas or span gas spills out of the probe when you are doing
zero or span checks.)
(4) You do not need a heated CLD to test spark-ignition engines.
(b) Using CLDs is generally acceptable even though they measure only NO and
NO